Model BBV-1000

General

A laboratory bench-top system for automatically determining annealing point temperature, strain point temperature or viscosity (Log poise 14.0 - 9.0) of a solid glass beam, rod, or tube below 1000C.

Operated according to ASTM C598, ASTM C1350M or modified versions of those test procedures

Includes

  • Bench top instrument console
  • 1000°C furnace
  • programmable process controller
  • measuring head with fused quartz sample pedestal (50mm diameter) and fused quartz load rod
  • type "S thermocouple
  • load weights
  • computer software
  • accessories and instruction manual. (Computer not included.)

Specifications

Model BBV-1000
Warranty 12 Months
Range Log 14.0 - Log 9.0 Poises
Atmosphere Air
Temperature Range RT to 1,000°C
Furnace Tube style
LVDT Displacement Range ±0.500 inch (±12.7 mm)
LVDT Sensitivity 0.7 mV/V/.001 inch(28 mV/V/mm)
Mass Set Slotted weight set 1-500 Grams
Thermocouple Type "S
Sample Size 2.53 to 3.78mm dia. Rod (or) 2.21 to 3.31mm bar
Power Requirements 120vac/10amp or 240vac/5amp, 50 or 60 Hz
Computer Requirements Desktop with available PCI slot
Power Control SSR
Pedestal Fused Quartz - 50mm diameter
Load rod Fused Quartz Shepherds hook
Furnace Movement Manual lift
Temperature Control Honeywell UDC 3200 Process controller
Computer Interface Software Included
25mm Diameter Pedestal Optional
Bench-top Footprint (stand)Length x Depth x Height 12 x 14 x 33 (31 x 36 x 84 cm)
Bench-top Footprint (controller)Length x Depth x Height 18 x 12 x 5 (46 x 31 x 13 cm)

View Products by Category:

 

Search
clear

FROM THE BLOG...

Measuring Flexural Creep of Ceramic Investment Casting Shells

Measuring Flexural Creep of Ceramic Investment Casting Shells Understanding the creep behavior...

5 Ceramic Products That Changed the World

5 Ceramic Products That Changed the World!   When you think of ceramics, many people may flash...

Orton Welcomes New General Manager!

The Board of Trustees of the Edward Orton Jr. Ceramic Foundation announces the retirement of its General...