Model ANS-1000

General

A laboratory, bench-top system for automatically determining annealing point temperature and strain point temperature of a glass fiber below 1000C.

Operated according to ASTM C-336 Method A or modified version of this procedure.

Includes

  • Bench top instrument console
  • 1000°C furnace
  • programmable process controller
  • measuring head with LVDT sensor
  • type "S" thermocouple
  • 1Kg load weight
  • computer software
  • accessories and instruction manual. (Computer not included.)

Specifications

Model ANS-1000
Warranty 12 Months
Accuracy +/-2C
Atmosphere Air
Temperature Range RT to 1000°C
Furnace Core Inconel
Heated Length 10.75 inches (273 mm)
LVDT Displacement Range ±0.500 inch (±12.7 mm)
LVDT Sensitivity 0.7 mV/V/.001 inch(28 mV/V/mm)
Load 1 Kg
Thermocouple Type "S
Sample Size 508 mm long x 0.65 mm diameter
Power Requirements 120vac/10amp or 240vac/5amp, 50 or 60 Hz
Computer Requirements Desktop with available PCI slot
Power Control SSR
Temperature Control Honeywell UDC 3200 Process controller
Computer Interface Software Included
SRM Calibration samples Optional
Bench-top Footprint (stand)Length x Depth x Height 12 x 12 x 33 (31 x 31 x 84 mm)
Bench-top Footprint (controller)Length x Depth x Height 18 x 12 x 5 (46 x 31 x 13 cm)

View Products by Category:

 

Search
clear

FROM THE BLOG...

TempTab Brochure for the Dental Industry Announced

Verifying temperatures in your Sintering Furnace?  The August Buyers Guide of LMT magazine offers TempTab...

High Temperature Ceramics for Extreme Environments Workshop - June 18 & 19, 2019

The Ohio State University College of Engineering Department of Materials Science and Engineering...

2019 Introduction to Refractories Short Course

Orton's 2019 "Introduction to Refractories" short course is coming up in August. This course is an intensive...