Model ANS-800

General

A laboratory bench-top system for automatically determining annealing point temperature and strain point temperature of a glass fiber below 800C.

Operated according to ASTM C-336 Method A or modified version of this procedure.

Includes

  • Bench top instrument console
  • 800°C furnace
  • programmable process controller
  • measuring head with LVDT sensor
  • type "S" thermocouple
  • 1Kg load weight
  • computer software
  • accessories and instruction manual. (Computer not included.)

Specifications

Model ANS-800
Warranty 12 Months
Accuracy +/-2C
Atmosphere Air
Temperature Range RT to 800°C
Furnace Core Copper
Heated Length 10.75 inches (273 mm)
LVDT Displacement Range ±0.500 inch (±12.7 mm)
LVDT Sensitivity 0.7 mV/V/.001 inch(28 mV/V/mm)
Load 1 Kg
Thermocouple Type "S
Sample Size 508 mm long x 0.65 mm diameter
Power Requirements 120vac/10 amp or 240vac/5amp, 50 or 60hz
Computer Requirements Desktop with available PCI slot
Power Control SSR
Temperature Control Honeywell UDC 3200 Process controller
Computer Interface Software Included
SRM Calibration samples Optional
Bench-top Footprint (stand)Length x Depth x Height 12 x 12 x 33 (31 x 31 x 84 mm)
Bench-top Footprint (controller)Length x Depth x Height 18 x 12 x 5 (46 x 31 x 13 cm)

View Products by Category:

 

Search
clear

FROM THE BLOG...

Orton Welcomes New General Manager!

The Board of Trustees of the Edward Orton Jr. Ceramic Foundation announces the retirement of its General...

How Kiln Environment will affect a Pyrometric Cone

Those experienced in firing ceramics know that kiln atmosphere can have a major influence on the reaction...

TempTabs: What They Are & How They Work - Watch This Video!

TempTabs are used to measure temperature uniformity, consistency and repeatability in high temperature...