General

​

Orton dilatometer Model 2010C is a horizontal, digital dilatometer that measures the expansion or contraction of a single test specimen from -190°C to +500°C or RT to 1000°C with on-board, stand-alone, programmable process controller and linear motion measuring head assembly equipped with LVDT sensor.

​

Operated according to ASTM E-228, ASTM C-372, or ISO 7991

​

Includes

​

  • Bench top instrument console

  • 1000°C split tube furnace

  • Programmable process controller

  • Fused Quartz sample holder and Fused Quartz probe rod assembly

  • Liquid Nitrogen delivery system

  • Controlled Atmosphere components

  • Type "N" thermocouple

  • 1 (25.4mm) calibration samples

  • Computer software

  • Accessories and instruction manual. (Computer not included)

​

​

Specifications

 

Model: DIL 2010 C

Warranty: 12 Months

Temperature Range: RT to 1,000°C or -180C to 500C

Furnace: Split Tube style, 12 inch heated zone

Thermocouple: Type "N" 2 wire

Sample Holder and Probe Rod: Fused Quartz

Sample Size (max): 50 mm long x 10 mm round or square

LVDT Displacement Range: ±0.100 inch (±2.54 mm)

Displacement Resolution: 0.0000009 inch or 0.9 micro-inch (0.00002 millimeter or 0.02 microns)

PLC Resolution for a 1 Sample: 0.00009%

Reproducibility Range: ± 0.004 Percent Linear Change

Contact Load: Adjustable - minimum 4 grams

Temperature Control: User Programmable, 20-segment, PID Controller with Melting Point Protection

Heating Rate: 1 to 30°C/minute, programmable at 0.01°C increments

Power Requirements: 120vac/10amp or 240vac/5amp 50/60Hz

Data Acquisition: Onboard Process controller - data stored at 1°C increments, real-time data collection or data retrieval to computer

Data Analysis: Remote programming and analysis Software (Requires English Language Version of Windows OS)

Computer Interface: RS232 or USB with adapter

Factory Calibration: high purity platinum

Calibration Sample: 99.8 Alumina; Platinum available as Option

Water Cooled Bulkhead: Optional (water circulation system not included)

Measuring Head Cover: Included

Controlled Atmosphere Option: Included

Fused Silica Calibration Sample: Optional

Bench-top Footprint; Length x Depth x Height: 49 x 14 x 24 (1,250 x 360 x 600 mm)

​

​

​