Model ANS-800​
​
​
​
​
​
​
​
General
​
A laboratory bench-top system for automatically determining annealing point temperature and strain point temperature of a glass fiber below 800C.
​
Operated according to ASTM C-336 Method A or modified version of this procedure.
​
Includes
​
-
Bench top instrument console
-
800°C furnace
-
Programmable process controller
-
Measuring head with LVDT sensor
-
Type "S" thermocouple
-
1Kg load weight
-
Computer software
-
Accessories and instruction manual. (Computer not included.)
​
​
Specifications
Model: ANS-800
Warranty: 12 Months
Accuracy: +/-2C
Atmosphere: Air
Temperature Range: RT to 800°C
Furnace Core: Copper
Heated Length: 10.75 inches (273 mm)
LVDT Displacement Range: ±0.500 inch (±12.7 mm)
LVDT Sensitivity: 0.7 mV/V/.001 inch(28 mV/V/mm)
Load: 1 Kg
Thermocouple: Type "S"
Sample Size: 508 mm long x 0.65 mm diameter
Power Control: SSR
Power Requirements: 120vac/10amp or 240vac/5amp, 50/60 Hz
Computer Requirements: Windows OS
Temperature Control: Honeywell UDC 3200 Process controller
Computer Interface Software: Included
SRM Calibration samples: Optional
Bench-top Footprint (stand) Length x Depth x Height: 12 x 12 x 33 (31 x 31 x 84 mm)
Bench-top Footprint (controller) Length x Depth x Height: 18 x 12 x 5 (46 x 31 x 13 cm)
​
​